Need to be quickly call back ?

[contact-form-7 404 "Not Found"]

ME520 - Converter of signals and conditioner of analog sensors for strain gauges, DIN rail format.

APPLICATIONS

  • Conditioning of any kind of analog signal coming from a sensor to fit with data acquisition systems or Programmable Logic Controller (PLC)
  • Electrical and physical measurement: strength, torque, pressure, displacement, acceleration, strain gauges…, on test bench or control of industrial process.

DESCRIPTION

The single channel measurement model ME520 provides very complete analog outputs for measurement, conditioning and conversion of signals such as strength, torque, pressure, displacement, acceleration sensors or strain gauges.
Three parameters can be adjusted with an easy procedure that allows output trimming depending on the sensor:
  • A potentiometer to trim gain
  • A potentiometer to trim zero
  • A push button to trim calibration
A wide range of options is available: isolated power supply and many other sensors adapted to ME520 or ME52C.

TECHNICAL FEATURES

Isolated General Supply24 VDC (run from 18 to 36 VDC).
Operating temperature range-10 to +60°C
Storage temperature range-40°C to +70°C
Dimensions of rail DIN unitH: 99 L: 17.5 P: 112 mm
Connectors4 x 3 limits
Approx. weight (without option)About 110 grams
Input1 sensor or 4x 350 Ω sensors set up in bridge, trimmed together an supplied with 5Vdc max
PrecisionJumper switch configuration: 0-10V ; ±10V, 0-20mA ; 4-20mA or ±20mA
Bandwidth0.05%
Calibration trimJumper switch configuration: 2 KHz or 20 KHz at 3dB
Full bridgeJumper switch configuration: 2 KHz or 20 KHz at 3dB
Excitation bridge0.1 to 10mV/V
Zero adjustment120 to 10K Ω
Gain adjustment5 or 10V
Sensitivityyes
yes
0.1 to 30mV/V

OPTIONS

ME26-FLT Filtered outputButterworth 6th fixed order (BP on request)
ME26ISO 3 channels galvanic isolationInput / output: 400 V – Input / power supply: 1000 V

VARIATION

Converter voltage/ voltage; voltage / current; current / current; current / voltage
  • Voltage input: all values from ±50mV to ±100V
  • Current input: all values from ±0.1mA to ±100mA
  • Voltage output: all values from 0 to 10V or ±10V
  • Current output: all values from 0 to 20mA or ±20mA

ME520 – U/I, I/U, I/I or U/U module

Amplifier conditioner for standard strain gauge (force, torque, pressure…)
  • Full bridge from 120 to 10k Ω
  • Possibility to set up 4 x 350Ω sensors in bridge, trimmed together and supplied with 5Vdc max
  • Bridge excitation 5 or 10V
  • Sensitivity from 0.1 to 30mV/V
  • Calibration trim from 1 to 100mV/V

ME520-AJ module

Amplifier conditioner for piezo-resistive sensors
  • Full bridge from 120 to 10k Ω
  • Possibility to plug 4 x 350Ω sensors in parallel, aligned in sensitivity and up=5Vdc
  • Bridge excitation 5 or 10V
  • Sensitivity from 0.1 to 30mV/V
  • Calibration trim from 1 to 100mV/V

ME520-AS module

Amplifier conditioner for strain gauge (extensometry)
  • 4-wire full bridge stain gauge
  • Complement of ¼ and ½ bridge 350Ω (120Ω on demand)
  • -Bridge supply: 5V or 10V for 4-wire
  • Sensitivity adjustable from 0.1mV/V to 30mV/V
  • Calibration trim from 0.1 to 10mV/V (depending on ranges)

ME520-AJ Microstrain module

Amplifier conditioner with high zero shift (2.5V) for integrated electronic sensors (such as some inclination or acceleration sensors)
  • 3-wire sensor input: ranges 2.5V ±100mV at 30VDC
  • Sensor supply 5V or 10V (other values on demand)
  • 2-wire input (BNC connector)
  • Constant current supply (7mA), unloaded voltage 24VDC
  • Adjustable input voltage from 100mV to 30V
  • Trim button (depending on range) from 100mV to 10V
  • Bandwidth: 1KHz to 2KHz or 20KHz at -3dB

ME520-ACE/ INC module

Amplifier conditioner for PT100 probe (platinum probe 100? at 0°C)
  • 2, 3 or 4-wire probe
  • Sensor supply 5V or 10V (other values on demand)
  • Sensor supply: 1mA
  • Various input ranges

ME520- PT100 module

For other enquiries, please contact us.